Complete Guide to Dipslides: How to Use, Incubate and Read Results
Learning how to use dipslides is straightforward: you expose the agar to a liquid or surface sample, return the slide to its tube, incubate it under the conditions specified for that product code, then compare the growth against the comparison charts supplied with it. The whole process is designed to be carried out on site, by technical and non-technical users alike.
What it gives you is a routine check you can repeat and trend over time. It is not a substitute for confirmatory laboratory analysis. As our instruction leaflet sets out, the comparison charts are for presumptive analysis, and confirmation of target organisms by laboratory methods is advised. The media, the incubation conditions and the way you read the result all depend on the product you have chosen and what you are monitoring.
What Does a Dipslide Test For?
A dipslide is a plastic paddle coated on both sides with prepared culture media, held inside a sealed tube. When the agar contacts a liquid or a surface, any microorganisms present are transferred onto the media. Incubation then encourages those organisms to multiply until the growth is visible to the eye and can be compared against a chart.
Different media are formulated for different monitoring purposes, which is why the range runs to dozens of product codes rather than a single universal slide. Depending on the product you select, our dipslides support checks for total count, yeasts and moulds, coliforms and E. coli, Enterobacteriaceae, Staphylococcus aureus and Listeria, among other specified target groups. Many slides are dual-media, with a different agar on each side, so that one sample can be assessed for two things at once.
This is the part worth getting right before anything else. A slide can only grow what its media is designed to support, so a result tells you about the target groups for that product and nothing more. It does not confirm the identity of an organism on appearance alone. If you are not certain which slide suits your sample or your target organism, browse the general purpose range, check our frequently asked questions, or speak to our technical team before you sample rather than after.
What You Need Before You Start
- The correct dipslide for the monitoring task, in date and stored correctly.
- An incubator able to hold the temperature specified for that product.
- A means of labelling and recording the sample location, date, time and product code.
- The instruction leaflet, including the bacteria and yeast chart, the mould chart and the growth characteristics tables.
- Any PPE, sampling, hygiene and disposal procedures that apply at your site.
Inspect every slide before you open it. If there is any sign of growth on the agar before use, do not use it. Slides should be kept in a cool, dry place at an optimum of 8–15°C and used before the expiry date printed on the box.
How to Use Dipslides Step by Step
The sequence below is the core process. Read it through once before your first sample, then follow the surface or liquid method as appropriate.
- Check the product code, expiry date, storage condition and the appearance of the agar before opening.
- Label the tube, or prepare your sample record, before you take the sample.
- Remove the cap and withdraw the combined cap and tongue unit from the outer tube. Do not touch the culture gel at any point.
- Sample the surface or the liquid using the appropriate method described below.
- Allow excess liquid to drain where relevant, return the slide to its tube and push to seal firmly.
- Incubate the slide upright, at the time and temperature specified for that product code.
- After the full incubation period, compare the growth on both sides against the correct chart and record what you find.
- Handle and dispose of the used slide according to the published guidance and your own site procedure.
How to Sample a Surface
Hold the slide by the tabs at each end and press one side of the agar against the surface under test. The ribbed section above the agar is designed to bend, which allows the media to make full contact with the surface rather than touching at a single point.
Repeat with the second side on an area close to the first sampling point. Return the slide to its outer tube, push to seal firmly, label it and incubate.
When you are monitoring trends rather than taking a one-off reading, keep your contact area and technique consistent between samples. Two results are only comparable if they were taken the same way.
How to Sample a Liquid
There are two accepted methods, and which you use depends on how you can access the sample.
Method 1 — immersion. Dip the tongue into the fluid until the agar is fully submerged. Wait approximately ten seconds, then remove the slide and allow excess fluid to drain for a few seconds.
Method 2 — running sample. Hold the slide under a running liquid or spray sample so that both sides of the agar are covered. Allow excess fluid to drain for a few seconds.
With either method, return the slide to its tube, push to seal firmly, label it and incubate. Do not touch the gel or rest it against another surface while you are working.
Dipslide Incubation: Time, Temperature and Position
Slides should be incubated upright.
There is no single incubation time or temperature that applies across the range, and using one figure for every product is a common source of unreliable results. Our incubation table lists requirements by product code at 22°C, 30°C and 35°C, with periods running from 24–48 hours up to 24–120 hours.
Set BT2, BTM2 and PO side by side:
|
Product code |
Temperature |
Period |
|
BT2 |
35°C |
24–48 hours |
|
BTM2 |
30°C |
24–120 hours |
|
PO |
22°C |
24–120 hours |
Three different products, three different sets of conditions. Always check the table for the slide in front of you, or incubate at the temperature required for growth of your target organisms. Where formal guidance applies to your monitoring, for example HSE guidance, that guidance should be followed.
In practice this matters in both directions. Reading a slide before its stated period is complete risks missing growth that had not yet become visible. Equally, incubating at the wrong temperature, or extending the period without a documented reason, changes the conditions your result was produced under and makes it harder to compare against your previous samples.
How to Read and Interpret Dipslide Results
Once the full incubation period has elapsed, examine both sides of the slide and note the amount, colour and pattern of visible growth. On a dual-media slide the two sides are answering different questions, so record them separately.
For a quantitative reading, compare the slide against the chart supplied and take the result from the closest matching picture. There are separate charts for bacteria and yeasts and for moulds, so make sure you are using the right one. Liquid results are read as colony-forming units per millilitre (CFU/ml) and surface results as colony forming units per square centimetre (CFU/cm²).
It is good practice to compare growth against an unused slide as a negative control. Heavy contamination does not always look dramatic. Where growth is dense and evenly spread, it can present as a subtle film across the whole agar surface, which is easy to underestimate without a clean slide alongside it for comparison.
Colour and colony descriptions in the growth characteristics tables are indicators, not identifications. Those descriptions are based on ATCC control strains and there is natural variation between organisms, with different species sometimes producing a similar appearance. Treat the reading as presumptive. Where the result matters, confirm it through laboratory methods.
One thing to avoid is turning a chart reading into a universal clean or dirty verdict. There is no single acceptable level that applies everywhere. What counts as acceptable depends on your site, your process, your monitoring plan and any guidance that applies to your industry, so a result should always be assessed against those rather than against a general benchmark.
How to Record and Use the Result
For each sample, record the date, the sampling point, the product code, the incubation time and temperature, the result for each side of the slide, and anything unusual you noticed.
The value of routine dipslide monitoring builds up over time. Used correctly, dipslides give reliable trend results, and a consistent record lets you see whether microbial levels at a given point are stable, rising or falling. A single reading in isolation tells you much less than the same test repeated the same way over several weeks.
If a result is unexpected, unusually high, or suggests the presence of a target organism, follow your organisation's escalation procedure and consider confirmatory laboratory analysis. Retest frequency and corrective action should come from your monitoring plan or applicable guidance rather than from the slide itself.
Common Mistakes to Avoid
- Selecting a slide whose media does not suit the organism or sample being monitored.
- Using an expired slide, or one showing growth before it was opened.
- Touching the agar, or letting it contact a surface you did not intend to sample.
- Failing to label the slide or record the sampling point.
- Applying one incubation time and temperature across every product code.
- Reading the slide before the stated incubation period has finished.
- Comparing against the wrong chart, or treating the closest image as an exact laboratory count.
- Treating presumptive growth as a confirmed identification.
- Disposing of a contaminated slide without following the published method and your site procedure.
Storage and Safe Disposal
Store slides in a cool, dry place at an optimum of 8–15°C and use them before the expiry date on the box.
Rapid fluctuations in temperature can cause condensation to form in the bottom of the tube and can dehydrate the agar. This does not affect performance, provided the agar still fills the tray. Inspect slides before every use, and do not use one showing growth before opening, as it may contaminate your sample.
Used and contaminated slides should be incinerated, autoclaved, or soaked in disinfectant for a minimum of 24 hours before disposal. Local regulations and your own site waste procedures apply alongside this.
When to Ask for Technical or Laboratory Support
- Contact our technical team, or arrange laboratory analysis, when:
- You are unsure which dipslide suits your sample or target organism.
- Growth does not resemble the images or descriptions supplied.
- Different organisms could plausibly account for what you are seeing.
- A result carries safety, compliance or process implications.
- Your monitoring plan calls for definitive identification or a fully quantitative laboratory result.
You can reach the team through our contact page.
Choose the Right Slide for the Job
Browse our dipslide range and select the product that matches your sample type and target organisms, including our food and water and industrial ranges.
Get the Detail Before you Sample
Download the instruction leaflet for the full incubation table, growth characteristics and comparison charts, read more about how our dipslides are made and used, or contact our technical team for help choosing a slide.
Frequently Asked Questions
How long should dipslides be incubated, and at what temperature?
It varies by product code. Our incubation table lists periods from 24–48 hours up to 24–120 hours, at 22°C, 30°C or 35°C. Check the table for the slide in front of you, or incubate at the temperature and period required for your target organisms. Where formal guidance such as HSE guidance applies to your monitoring, follow that guidance.
What do CFU/ml and CFU/cm² mean on a dipslide chart?
Colony-forming units per millilitre and per square centimetre. CFU/ml is used for liquid samples and CFU/cm² for surface samples, which is why the comparison chart carries a separate scale for each.
Is there an acceptable level of growth on a dipslide?
There is no universal pass or fail figure. What counts as acceptable depends on the site, the process, the monitoring plan and any guidance that applies to your industry. The chart tells you the approximate level present; your monitoring plan tells you what to do about it.
What does no growth after incubation mean?
Check the full period for that product code first, because some slides run to 120 hours and reading at 48 hours may simply be too early. Where no growth is visible after the complete period, that indicates no detectable growth of the groups the slide's media is designed to support. It says nothing about organisms that the media does not support, which is why slide selection matters.
Can a dipslide identify a specific bacterium?
No. The growth characteristics and agar colours in our tables are based on ATCC control strains and act as a guide only, and different organisms can produce a similar appearance. Dipslide results are presumptive, and confirmation of target organisms through laboratory methods is advised.
How should used dipslides be disposed of?
Incinerate them, autoclave them, or soak them in disinfectant for a minimum of 24 hours before disposal. Your own site waste procedures and any local regulations apply alongside this.